Published July 8, 2026 at 1:35 PM
Salinity measurement has historically been the domain of the hydrometer, the simple floating glass cylinder that sits higher or lower depending on the density of the liquid. The method works but has practical limitations: it needs a large sample, is slow to stabilize and is temperature-dependent. For modern aquaculture, aquarium and marine biology applications, handheld refractometers have almost entirely replaced the hydrometer. An analog refractometer with automatic temperature compensation is the budget solution that delivers professional-quality measurement for hobby and small-scale professional use.
What the Extech RF20 is
Salinity refractometer (0 to 100 ppt) with temperature compensation (ATC) (RF20) is an analog handheld refractometer from Extech for measuring salinity. The instrument:
- Measures salinity from 0 to 100 ppt (parts per thousand).
- Has automatic temperature compensation (ATC) from 10 to 30 °C.
- Uses optical refraction through a prism and lens.
- Has an easy-to-read scale with typical graduations of 1 ppt.
- Requires only 2-3 drops of sample.
- Has an adjustable prism and lens.
- Comes with a carrying case, calibration screwdriver, plastic pipette and calibration solution.
- Is a budget solution for semi-professional and hobby use.
Basic specifications:
- Manufacturer: Extech (USA)
- Model: RF20
- Salinity measuring range: 0-100 ppt (parts per thousand)
- Resolution: 1 ppt
- Temperature compensation: Automatic (ATC) 10-30 °C
- Sample volume: 2-3 drops
- Measuring time: ~3 seconds
- Calibration: Single-point with pure water or calibration solution
- Type: Analog optical (traditional design)
- Accessories: Carrying case, calibration screwdriver, plastic pipette, calibration solution
Salinity units and conversion
Salinity is expressed in several different units:
- ppt (parts per thousand): 1 ppt = 1 g of salt per 1,000 g of water = 0.1%.
- PSU (Practical Salinity Unit): Almost identical to ppt for normal seawater composition.
- Salinity percentage (S%): S% = ppt / 10. Seawater is typically 3.5% = 35 ppt.
- Specific gravity (SG): Weight of a solution ÷ weight of the same volume of pure water. Seawater at 35 ppt has an SG of 1.026.
- Refractive index (n): An optical parameter. Pure water n = 1.333. Seawater n ≈ 1.339.
Refractometric measurement gives the refractive index directly; the scale converts it to ppt using a calibration curve for NaCl solution.
Natural seawater as a reference
Global seawater salinity varies geographically:
- Atlantic Ocean: 34-36 ppt (average ~35 ppt).
- Pacific Ocean: 33-35 ppt.
- Red Sea: 38-40 ppt (high evaporation).
- Baltic Sea: 5-10 ppt (brackish water).
- White Sea: 20-25 ppt.
- Dead Sea: 300+ ppt (extremely saline).
- Caspian Sea: 12-13 ppt.
- Mediterranean Sea: 36-40 ppt.
For aquarists with reef (coral) aquariums, the optimal salinity is 34-35 ppt. Deviations of more than 3 ppt stress invertebrates; deviations of more than 5 ppt cause death.
Osmotic stress and salinity tolerance
The salt concentration in water directly affects the osmotic balance of aquatic organisms:
Stenohaline (narrow salinity tolerance): Marine fish (cod, tuna), corals, most invertebrates. Tolerance range is typically ±2 ppt around natural seawater.
Euryhaline (wide salinity tolerance): Salmon, eel, mollies, Artemia. Tolerance range is 5-30 ppt or more.
Physiological consequences of salt stress:
- Salinity too high: Water is drawn out of the cells (dehydration). Fish become lethargic and corals close up.
- Salinity too low: Water flows into the cells (swelling). Cell membranes can rupture. Corals collapse.
In the event of an acute change in salinity (shock), sensitive species can die within hours. Slow acclimation (0.5-1 ppt per day) is necessary for fish to survive a change in salinity.
Calibration procedure
Single-point calibration with calibration solution (typically 35 ppt standard):
- Clean the prism surface with pure water and a soft cloth.
- Place 2-3 drops of calibration solution on the prism.
- Close the cover plate.
- Point the refractometer toward a light source (daylight works best).
- Read the value.
- If the value is not 35 ppt, adjust it with the calibration screwdriver via the calibration screw.
- Repeat until the value is correct.
- Rinse the prism with pure water and dry it.
Calibration should be checked monthly or before critical measurements.
How the ATC system works
The refractive index of a salt solution changes with temperature. At 20 °C, 35 ppt has n = 1.3395; at 30 °C the same solution has n = 1.3403 (a difference of 0.0008, which corresponds to a reading error of ~2 ppt).
The ATC (Automatic Temperature Compensation) in the analog RF20 uses a bimetallic construction inside the optics. The bimetal bends with temperature and mechanically compensates for the change in refractive index. Advantage: it works without electronics. Disadvantage: it is limited to a 10-30 °C range (outside this range the bimetal becomes non-linear).
Practical consequence: samples from cold water (refrigerator, ice water) should be warmed to room temperature before measurement for the ATC to work correctly.
Where a salinity refractometer makes a real difference
Saltwater and coral aquariums (hobby): Weekly verification of salt concentration for the health of fish and corals.
Commercial aquaculture: Marine fish farms (salmon, trout). Brackish-water farms (eel, giant prawns). Salt concentration affects growth rate and disease incidence.
Ocean research and marine biology: Field measurement of local salinity variations.
Salt ponds and solutions: Salt production ponds where the concentration must be monitored.
Preservation and salting: Food production where brine concentration is critical for preservation.
Winter road maintenance: Salt solution concentration for spreading products. NaCl solution above 22% freezes at -18 °C.
Health and beauty industry: Salt scrubs and salt solutions for therapeutic use.
Agriculture and horticulture: Verification of irrigation water. High salt levels damage crops.
Swimming pools with salt chlorination systems: Salt concentration is typically 3-4 g/L (3-4 ppt). Refractometric verification.
Education in aquatic biology and chemistry: Practical demonstration.
Analog vs. digital refractometer
Analog RF20 vs. digital Palm Abbe or ATAGO SugarMaster:
Analog RF20 advantages:
- Low cost (1,500-2,500 SEK).
- No battery.
- Mechanically robust.
- Quick to use outside a lab environment.
Analog RF20 disadvantages:
- Requires daylight or a strong lamp for reading.
- Operator reading errors are possible.
- No data storage or documentation.
- Fixed calibration point.
Digital Palm Abbe advantages: Direct reading on an LCD, temperature compensation outside 10-30 °C, data storage possible, elimination of operator error.
Choice: analog for hobby and budget applications; digital for professional documentation requirements.
What you get for your money
Extech RF20 analog salinity refractometer, measuring range 0-100 ppt (parts per thousand), resolution 1 ppt, automatic temperature compensation (ATC) 10-30 °C via a built-in bimetallic construction, optical refraction through a prism and lens, sample volume of just 2-3 drops, measuring time ~3 seconds, adjustable prism and lens, single-point calibration, complete with carrying case, calibration screwdriver, plastic pipette and calibration solution.
SEK 1,607 is the investment in the instrument. Compared with a hydrometer with a thermometer (200-500 SEK, requires 100+ ml of sample, slow measurement, difficult to calibrate), refractometric measurement is considerably faster and more convenient. Compared with a digital Palm Abbe salinity refractometer (5,000-7,000 SEK, digital reading and 0-50 °C ATC), the RF20 is a budget solution suited to hobby and semi-professional use.
Read more: Extech RF20 analog salinity refractometer in the shop →